SE Lamidesk
LexJetAU 143x118v3
Sign China 2
1389_0816_Vinyl_wrap_pull_143_x_118px_fa.jpg
New ad in orange
AGS logo for Wideformat 120724
animation 2
1601 GJS website
 
 
 
 
Join our newsletter mailing list
Email:

How to measure printed colour on glass, plexiglass and other transparent media 

An image printed onto glass is viewed through an illumination from behind the image. Measurement with traditional reflective spectrophotometers is therefore not applicable.

This document describes how the printed colours can be measured with a transmission spectrophotometer in order to calibrate and profile the digital printer correctly for such applications.

This document also covers the colour measurement techniques of ink on all transparent media.

This document assumes a basic understanding of colour management in wide format printing and the rule of a spectrophotometer as a measuring device for building calibration curves and ICC profiles. We have used the Barbieri Spectro LFP RT for this exercise.

Media thickness
The Spectro LFP RT can measure thin, flexible transparent or translucent materials.

Measuring thicker material such as glass however does require a basic understanding on the limitations.

The picture below shows the optical situation:

the_optical_situation.jpgThe measuring head touches the glass and is in a distance of the glass thickness to the lamp/ diffusor. In this situation, the measuring head can see also light coming from outside and therefore cause wrong readings. This error does affect only dark colors.

To avoid measuring errors caused by straylight, a thickness up to 5 mm is acceptable, 3 mm recommended.

Patch size
The targets to be measured should have patches which are at least 3 times larger than the measuring aperture used. This is necessary to avoid influence of neighbor patches.

For ICC profile creation, we recommend the use of the 2mm measuring aperture of the Spectro LFP RT and the target: Barbieri CMYK_02 available for download at http://www.profile-xpert.com/.

Printing considerations
Basic rule: Glass should be measured in the same way as the final result will be looked at. The diffusor of the instrument (light source for transmissive readings) should be at the side of the light source and the measuring head at the side of view.

This can cause 2 situations:

1) Print side towards measuring head: this is the preffered measuring method, as the measuring head only sees the light coming through the patch to be measured.

print_side_towards_measuring_head.jpg2) Print side behind (towards diffusor): the thicker the glass, the more light can come sideways into the measuring head causing erroneously dark patches to become lighter.

Preparation

 capture.jpg patch_size.jpg 

After printing, cut the target on the indicated lines. Glass must be cut neatly.

Use sample holder model 2 supplied with your instrument. Attention to not scratch the measuring table (and diffusor). Fix the glass to the sample holder using a tape. Fix the glass on the top end of the sample holder

Reference area
Barbieri targets have a “reference area” in the upper left corner of the target. Do not cut off this reference area.
If printing with white ink backing, also the reference area must be printed. capture_4.jpg

Measurement
Measurement is executed automatically in “fast” measuring mode in the same way as measuring standard materials.

Opal glass
capture_5.jpgOpal glass is measured in the same way as fully transparent glass is measured. Measuring time is slower, as the instrument gets lower light level and therefore adjusts measuring time accordingly. In this lower light condition, the stray light influence is more critical and the thinner the Opal glass is, the better the measuring results.

Coloured glass
All measurements described in this document are relative to the “reference area” on the target. This means, the instrument assumes the white point to be at the reference area and gives it a value of L*=100, a*=0, b*=0.

This works great for colour management applications which assume a “white background”, as it corresponds to the interpretation of the human eye which also adapts image interpretation to this reference white point.

It must be noted, that when using coloured glass, this method makes it impossible for the ICC profile to use the “absolute colourimetric intent” to match colours as “absolute” and “relative” intents are equal.

If absolute colourimetric matching is required, the reference area on the target must be substituted with a transparent area of same thickness.

Spot measurement

 capture_6.jpg
 capture_7.jpg

Single measurement of coloured patches is supported by the measuring software Profile-Xpert Gateway. Please refer to the operating manual of this software for detailed information.

When measuring a glass sample, please note that the instrument needs first to be calibrated on a clear area and then continue spot measurement on the patch to be measured:

Transmission vs. reflection
capture_8.jpgIn some applications, pictures are printed with a white backing. These pictures have the advantage, that they have a defined background colour and can be viewed also under daylight.

This raises the question: should it be measured/ profiled using transmission or reflection measuring mode?

The possible answers are as follows:
- If the picture is viewed in a dark room through a light table, transmission measurement is the correct answer
- If the picture is viewed in a bright room or under daylight, reflection measurement is the correct answer
- If the picture is viewed during the day under daylight, and during the night under backlit, a mixture of both should be used. The mixture is normally achieved by using the transmission method (giving normally higher saturated values) and printing the picture with lower saturation.

Colour space comparison between reflective measurement (green) and transmissive measurement (blue):

Reflection measurement
capture_9.jpgAs long as the printed side is towards the measuring head, reflection measurement on glass is done regularly same as measuring paper.
If the printed side is behind the glass, towards the diffuser, reflection measurement is not possible. The reason is, that the light is not reflected back into the optics.

Barbieri Electronics
www.BARBIERIelectronic.com

cielab
www.cielab.com.au

 

 

 

ROJO main FA4
PHD advert 03 19 17 v2
WFOL Academy Tile
jetmark sponsor ad 143x118
0977 optimus animated ad
Conect WFOL
 celmac
WFOL GIFs animated revised
 

Follow us on
 Twitter, Facebook and LinkedIn and see all our newsletters.
Twitter logo

Facebook logo

Linkedin logo

 

See all our videos on YouTube
YouTube logo